Development and testing of a neonatal head manikin with anatomically accurate upper airway for respiratory equipment testing and medical personnel training
https://doi.org/10.17021/1992-6499-2026-2-109-120
Abstract
The relevance of the study is determined by the lack of inexpensive and realistic models of newborn heads in the Russian Federation for testing non-invasive respiratory support equipment and training medical personnel. Most of the manikins available on the market either have insufficient anatomical accuracy and the impossibility of taking measurements, or are extremely expensive. Objective. development and testing of a neonatal head manikin with anatomically accurate upper airway for respiratory equipment testing and medical personnel training. Materials and methods. The study is based on the use of anonymized data from computed tomography of newborns without pathologies, obtained at the State Budgetary Healthcare Institution of the Sverdlovsk Region, the Yekaterinburg Clinical Perinatal Center. Based on the data, a 3D model of the head with realistic airways was created, reproducing the anatomical features. The materials included PLA plastic and two-component silicone. The tests were conducted using the CPAP Infant Flow device and high-precision measuring devices. The study period included the development, production and testing of the model over a period of 12 months. Results. The developed head model demonstrated high accuracy in reproducing airway pressure. Ultra-soft silicone allowed identifying areas of increased pressure on the skin, making the model useful for training personnel. The practical significance consisted in testing and adjusting respiratory equipment, as well as training medical personnel in conditions close to clinical ones. Conclusion. The developed manikin of the newborn head has proven its effectiveness as a tool for testing equipment and training personnel. However, further assessment of the model's durability and its adaptation for work with different types of respiratory systems is required, as well as further implementation of additional electronic measuring modules to expand the functionality.
Keywords
About the Authors
V. S. MakarovRussian Federation
Vsevolod S. Makarov - anesthesiologist-resuscitator of the Department, Yekaterinburg Clinical Perinatal Center; Junior Researcher of the Laboratory, Ural State Medical University.
Yekaterinburg
E. V. Shestak
Russian Federation
Evgenii V. Shestak - Cand. Sci. (Med.), Head of the Department, Anesthesiologist-Resuscitator of the Highest Qualification Category, Yekaterinburg Clinical Perinatal Center; Head of the Laboratory, Associate Professor of the Department, Ural State Medical University.
Yekaterinburg
V. Yu. Starkov
Russian Federation
Vadim Yu. Starkov - anesthesiologist-resuscitator of the Department, Yekaterinburg Clinical Perinatal Center; Junior Researcher of the Laboratory, Ural State Medical University.
Yekaterinburg
References
1. Newnam K. M., McGrath J. M., Estes T., Jallo N., Salyer J., Bass W. T. An integrative review of skin breakdown in the preterm infant associated with nasal continuous positive airway pressure. Journal of Obstetric, Gynecologic, and Neonatal Nursing. 2013; 42 (5): 508–516. doi: 10.1111/1552-6909.12233.
2. Boyar V. Pressure Injuries of the Nose and Columella in Preterm Neonates Receiving Noninvasive Ventilation via a Specialized Nasal Cannula: A Retrospective Comparison Cohort Study. Journal of Wound, Ostomy and Continence Nursing. 2020; 47 (2): 111–116. doi: 10.1097/WON.0000000000000616.
3. Naha N., Pournami F., Prabhakar J., Jain N. Nasal Injury with Continuous Positive Airway Pressure: Need for “Privileging” Nursing Staff. Indian Journal of Pediatrics. 2019; 86 (7): 595–598. doi: 10.1007/s12098-019-02960-1.
4. McCoskey L. Nursing Care Guidelines for prevention of nasal breakdown in neonates receiving nasal CPAP. Advances in Neonatal Care. 2008; 8 (2): 116–124. doi: 10.1097/01.ANC.0000317260.99072.ae.
5. Chae S., Shon S. Effectiveness of simulation-based interprofessional education on teamwork and communication skills in neonatal resuscitation. BMC Medical Education. 2024; 24 (1): 602. doi: 10.1186/s12909-024-05581-1.
6. Yousef N., Moreau R., Soghier L. Simulation in neonatal care: towards a change in traditional training? European Journal of Pediatrics. 2022; 181 (4): 1429–1436. doi: 10.1007/s00431-022-04373-3.
7. MacKinnon K., Marcellus L., Rivers J., Gordon C., Ryan M., Butcher D. Student and educator experiences of maternal-child simulation-based learning: a systematic review of qualitative evidence protocol. JBI Database of Systematic Reviews and Implementation Reports. 2015; 13 (1): 14–26. doi: 10.11124/jbisrir-2015-1694.
8. Shestak E. V. Respiratornye narusheniya u novorozhdennykh detey: novye tekhnologii ranney diagnostiki, lecheniya i profilaktiki lokal'nykh oslozhneniy = Respiratory disorders in newborns: New technologies for early diagnosis, treatment and prevention of local complications. Yekaterinburg; 2025: 276 p. (In Russ.).
9. Shestak E. V., Makarov V. S., Starkov V. Yu. Patent RF, no. 2852232, IPC G09B 23/28. Simulation model of a newborn's head and method of manufacture for testing non-invasive respiratory support and training medical specialists in working with respiratory equipment. Applicant and patent holder Ural State Medical University. Appl. no. 2024139057, filed 23.12.2024, publ. 05.12.2025, bull. no. 34. (In Russ.).
10. Buser T. J., Boyd O. F., Cortés Á., Donatelli C. M., Kolmann M. A., Luparell J. L., Pfeiffenberger J. A., Sidlauskas B. L., Summers A. P. The Natural Historian's Guide to the CT Galaxy: Step-by-Step Instructions for Preparing and Analyzing Computed Tomographic (CT) Data Using Cross-Platform, Open Access Software. Integrative organismal biology. 2020. Vol. 2, no. 1. P. obaa009. doi: 10.1093/iob/obaa009.
11. Sander I. M., McGoldrick M. T., Helms M. N., Betts A., van Avermaete A., Owers E., Doney E., Liepert T., Niebur G., Liepert D., Leevy W. M. Three-dimensional printing of X-ray computed tomography datasets with multiple materials using open-source data processing. Anatomical Sciences Education. 2017. Vol. 10, no. 4. P. 383–391. doi: 10.1002/ase.1682.
12. Weatherall A. D., Rogerson M. D., Quayle M. R., Cooper M. G., McMenamin P. G., Adams J. W. A Novel 3-Dimensional Printing Fabrication Approach for the Production of Pediatric Airway Models. Anesthesia and Analgesia. 2021; 133 (5): 1251–1259. doi: 10.1213/ANE.0000000000005260.
13. Kavanagh K. R., Cote V., Tsui Y., Kudernatsch S., Peterson D. R., Valdez T. A. Pediatric laryngeal simulator using 3D printed models: A novel technique. Laryngoscope. 2017; 127 (4): E132–E137. doi: 10.1002/lary.26326.
14. Eu D., Daly M. J., Taboni S., Sahovaler A., Gilbank A. N., Irish J. C. Evaluation of a 3D Printed Silicone Oral Cavity Cancer Model for Surgical Simulations. Journal of Personalized Medicine. 2024; 14 (5): 450. doi: 10.3390/jpm14050450.
15. Dundek M. L., Ng E. K., Brazil A. M., DiBlasi R. M., Poli J. A., Burke T. F. Evaluation of a Bubble CPAP System for Low Resource Settings. Respiratore Care. 2021; 66 (10): 1572–1581. doi: 10.4187/respcare.08948.
16. Kamath A. A., Kamath M. J., Ekici S., Stans A. S., Colby C. E., Matsumoto J. M., Wylam M. E. Workflow to develop 3D designed personalized neonatal CPAP masks using iPhone structured light facial scanning. 3D Print Med. 2022. Vol. 8, no. 1. P. 23. doi: 10.1186/s41205-022-00155-7.
17. Neonatal CPAP Mask Biomedical Engineering Senior Design Project 2022. University of Rochester. URL: https://www.hajim.rochester.edu/senior-design-day/neonatalcpap-adaptor/.
Review
For citations:
Makarov V.S., Shestak E.V., Starkov V.Yu. Development and testing of a neonatal head manikin with anatomically accurate upper airway for respiratory equipment testing and medical personnel training. Astrakhan medical journal. 2026;21(2):109-120. (In Russ.) https://doi.org/10.17021/1992-6499-2026-2-109-120
JATS XML